Overview

product_highlights
  • Measures up to 3 gases simultaneously under NDIR
  • Two auxiliary sensor inputs (electrochemical, paramagnetic, other)
  • Integration into OEM gas analyzer systems
  • Custom detectors and filters
  • Custom samples cells and sources
  • Custom materials and mechanical designs
product_overview

Infrared Industries makes available its industry-leading NDIR optical benches for custom uses, including integration into OEM gas analyzer systems. We offer a wide range of abilities and features including custom detectors and filters, custom samples cells and sources, as well as custom materials and mechanical designs. Infrared Industries was the initial supplier of benches to the top automotive manufacturers in the 1970s and has built upon that technology to offer you gas analyzer component products that fit a wide range of needs.

Need a Customized Application?

Do you have a unique application that consists of a different range of gas not currently detected in the IR-510? We can build a tailor-made optical bench that detects the gas of your choice.

Contact us today

The IR-510 NDIR Optical Bench offers greater design flexibility to OEMs due to its small overall size and low power consumption, combined with our ability to customize subsystem designs to your specifications. The enhanced optics and electronics dramatically reduce zero drift after the initial warm-up period. Temperature and pressure compensation eliminate the major causes of span drift found in many other NDIR instruments.

The IR-510 can fit a wide variety of enclosures. The software and communications protocol are designed to make system integration simple and fast. The command set has the flexibility to provide a variety of output configurations including pneumatic controls, printers, measurement data, and selected displays. The system monitors and corrects for critical operating parameters that affect performance and provides real-time status of the overall integrity and quality of the gas measurement.

Portable applications can be satisfied by populating the battery slot. The IR-510 interfaces with analog outputs and to a PC for complete control, display, and graphing capability. Recording of the output data within the bench itself allows the measurements to be replayed and analyzed at a later time.

The IR-510 also has a protected detector for accuracy and long life, integrated DC pump and solenoid drivers, graphic LCD driver, quick-clean sample cell design, and is DC powered and portable.

Specifications

General Specifications
Measuring method NDIR single beam
Electrochemical and paramagnetic also available
Gas measured User specified
Measuring range Per application
Data refresh rate 500 ms
Accuracy ± 1% of full scale
Repeatability ± 1% of full scale
Zero drift ± 1% of full scale per 24 hours
Span drift ± 1% of full scale per 24 hours
Linearity ± 0.5% of full scale fitted to theoretical curve
Noise level Less than 1% of full scale
Warm-up time Ready for use in less than 30 seconds; fully stabilized in 3 minutes
Temperature range 35° F to 110° F (2° C to 46° C)
Ambient humidity Up to 95% relative humidity
Compliance Meets or exceeds EN 14624
Outputs
Digital RS232
Host communication USB (adapter required)
Gas Flow System
Power consumption 2.9 watts average @ 12V DC
External power 10V - 16V DC, less than 1A
Materials Stainless steel sample cell, sapphire windows, selected o-rings, and Teflon tubing
Dimensions 6.3" x 3.5" x 1.2"
Weight 8 ounces (225 grams)
Standard Gases HC, CO, CO2, and O2; other gases also available
Accessories and Options Included Optional
One year standard warranty x
Extended warranty x
Analog: 1V, 5V, or 10V DC; or 4-20 mA x
Rechargeable lithium ion battery pack x
Kynar sample cell for corrosive sample streams x
InfraView Software for PC control x