IR-510 NDIR Optical Bench
- Overview
- Specifications
- Collateral
- Videos
- Industries
- News
- Parts
Overview
- Measures up to 3 gases simultaneously under NDIR
- Two auxiliary sensor inputs (electrochemical, paramagnetic, other)
- Integration into OEM gas analyzer systems
- Custom detectors and filters
- Custom samples cells and sources
- Custom materials and mechanical designs
Infrared Industries makes available its industry-leading NDIR optical benches for custom uses, including integration into OEM gas analyzer systems. We offer a wide range of abilities and features including custom detectors and filters, custom samples cells and sources, as well as custom materials and mechanical designs. Infrared Industries was the initial supplier of benches to the top automotive manufacturers in the 1970s and has built upon that technology to offer you gas analyzer component products that fit a wide range of needs.
The IR-510 NDIR Optical Bench offers greater design flexibility to OEMs due to its small overall size and low power consumption, combined with our ability to customize subsystem designs to your specifications. The enhanced optics and electronics dramatically reduce zero drift after the initial warm-up period. Temperature and pressure compensation eliminate the major causes of span drift found in many other NDIR instruments.
The IR-510 can fit a wide variety of enclosures. The software and communications protocol are designed to make system integration simple and fast. The command set has the flexibility to provide a variety of output configurations including pneumatic controls, printers, measurement data, and selected displays. The system monitors and corrects for critical operating parameters that affect performance and provides real-time status of the overall integrity and quality of the gas measurement.
Portable applications can be satisfied by populating the battery slot. The IR-510 interfaces with analog outputs and to a PC for complete control, display, and graphing capability. Recording of the output data within the bench itself allows the measurements to be replayed and analyzed at a later time.
The IR-510 also has a protected detector for accuracy and long life, integrated DC pump and solenoid drivers, graphic LCD driver, quick-clean sample cell design, and is DC powered and portable.
Specifications
General Specifications | |
---|---|
Measuring method | NDIR single beam
Electrochemical and paramagnetic also available |
Gas measured | User specified |
Measuring range | Per application |
Data refresh rate | 500 ms |
Accuracy | ± 1% of full scale |
Repeatability | ± 1% of full scale |
Zero drift | ± 1% of full scale per 24 hours |
Span drift | ± 1% of full scale per 24 hours |
Linearity | ± 0.5% of full scale fitted to theoretical curve |
Noise level | Less than 1% of full scale |
Warm-up time | Ready for use in less than 30 seconds; fully stabilized in 3 minutes |
Temperature range | 35° F to 110° F (2° C to 46° C) |
Ambient humidity | Up to 95% relative humidity |
Compliance | Meets or exceeds EN 14624 |
Outputs | |
Digital | RS232 |
Host communication | USB (adapter required) |
Gas Flow System | |
Power consumption | 2.9 watts average @ 12V DC |
External power | 10V - 16V DC, less than 1A |
Materials | Stainless steel sample cell, sapphire windows, selected o-rings, and Teflon tubing |
Dimensions | 6.3" x 3.5" x 1.2" |
Weight | 8 ounces (225 grams) |
Standard Gases | HC, CO, CO2, and O2; other gases also available |
Accessories and Options | Included | Optional |
---|---|---|
One year standard warranty | x | |
Extended warranty | x | |
Analog: 1V, 5V, or 10V DC; or 4-20 mA | x | |
Rechargeable lithium ion battery pack | x | |
Kynar sample cell for corrosive sample streams | x | |
InfraView Software for PC control | x |
Collateral
Industries
News
Parts
Template: single-product.php